Calibration Bath, Model 796 Orion Deep Immersion Liquid Bath
Our Price: Contact us for pricing

-40 °C to 300 °C (572°F)

New Available in two ranges, 976 M (-40 °C to 125 °C) and 796H (30 °C to 300 °C)
  • Stability of ±0.007 °C
  • Deep immersion depth (470mm or 18.5”) with low filling costs
  • Capable of calibrating liquid in glass thermometers, no level adapters required
  • Variable Speed Stirring - allowing optimum performance for a wide range of viscosities
  • Over and Under Temperature Protection - protect expensive Water Triple Point (WTP) Cells as well as over temperature conditions

  •  Orion Product Spec Sheet
    Calibration Bath, Model 915 Parallel Tube Liquid Bath
    Our Price: Contact us for pricing

    Large Volume Fluid Bath with Optional External Single and Two Stage Chillers
    As low as -65°C to 300°C
    OCEANUS 6 - P.O.T.T.S Temperature Calibrator
    Our Price: Contact us for pricing

    Block Calibrator or Automated Fixed Points

    -25°C to 140°C (282°F)
    10" Deep for Long Sensors
    Optional Inserts for Black Body & Liquid Bath

    • Universal Input Temperature Indicator included in S Model to allow for programming of up to five “correction” points
    • ITS-90 Fixed Points can be used with the Gallium and Triple Point of Water Fixed Point Cells
    • Dry or Liquid Bath
    • Windows Software and PC Interface
    • High Capacity Deep Block for Very Long Thermometers
    • Ideal for Laboratory Comparison and Fixed-Point Calibrations

    Calibration Baths, Hydra Series Oil Baths
    Our Price: Contact us for pricing

    New Range of Stirred Liquid Baths, Superior stability and vertical uniformity

    -80°C to 300°C (572°F)
    • Parallel Tube Action — Liquid flows up the rear volume of the bath and down the working volume. This action creates very small vertical and axial gradients. Enabling the smallest overall uncertainties
    • Heating — All heating is outside the container. By using a large area nickel foil heater the complete bath wall is heated uniformly
    • Cooling — The cooling is built-in and also surrounds the calibration volume creating a low temperature ambient in which the heater can function efficiently
    • Wider Temperature Range — A unique system removes the coolant from around the calibration volume above ambient and lets the bath heat up to 125°C (121°C is a key sterilization temperature)
    • Commercial Grade Chillers — The chillers are 1/3 horse power commercial grade units (not cheaper domestic grade as used by some manufacturers)
    • Stability — Better than ±0.01°C over the complete temperature range
    • Calibration Depth — Double the depth of some baths, the Hydra Range has up to 300mm depth of immersion
    • Circular Design — The circular design eliminates cold corners in tank-shaped baths
    • Fast Cool Down — Goes from ambient to -80°C in 150 min