Isotech offers NIST traceable Calibration services from -196 to 1200°C (2200°F) for a variety of instruments such as SPRT’s, RTD’s, Thermistors, Thermocouples, Readouts, Liquid-in-Glass Thermometers, Dry Block Calibrators and Electronic TC & RTD Simulators.

Our NIST Traceable Certificates always include data, a list of standards used and ITS-90 or CVD temp co-efficients where appropriate. Certificates are compliant with the format and guidelines outlined by ISO/IEC 17025:2005.

We utilize both; standard ITS-90 Fixed Point procedures and comparison techniques to offer a flexible solution depending on customer requirements. Our procedures are based on the technical information contained in NIST Technical Note 1265, “Guidelines for Realizing the International Temperature Scale of 1990 (ITS-90)” pertaining to secondary PRTs and in BIPM’s “Techniques for Approximating the International Temperature Scale of 1990.”

As the master distribution and service center for all Isotech and ASL equipment in North, Central, & South America, we know our equipment better than anyone. If by chance your equipment required service it can be performed at the same time without additional costs of shipping to another facility.

Best Measurement Uncertainties (k=2, 95% confidence)
 

Isotech Fixed Point Calibration Uncertainties

Parameter / Equipment

Range

Best Uncertainty (±)

Comment

SPRT / PRT / Digital Thermometers

-195.798°C

±8.2 mk (0.0082°C)

Liquid Nitrogen Comparison

SPRT / PRT / Digital Thermometers

-38.834°C

±3.4 mk (0.0034°C)

Mercury Triple Point Cell

SPRT / PRT / Digital Thermometers

0.01°C

±2.9 mk (0.0029°C)

Water Triple Point Cell

SPRT / PRT / Digital Thermometers

29.764°C

±3.0 mk (0.0030°C)

Gallium Triple Point Cell

SPRT / PRT / Digital Thermometers

156.598°C

±3.4 mk (0.0034°C)

Indium Fixed Point Cell

SPRT / PRT / Digital Thermometers

231.928°C

±3.4 mk (0.0034°C)

Tin Fixed Point Cell

SPRT / PRT / Digital Thermometers

419.527°C

±4.9 mk (0.0049°C)

Zinc Fixed Point Cell

SPRT / PRT / Digital Thermometers

660.323°C

±6.6 mk (0.0066°C)

Aluminum Fixed Point Cell