Home > Products >

QUICKCAL - Hand-Held Dry Block Temperature Calibrator
QUICKCAL - Hand-Held Dry Block Temperature Calibrator
 
Alternative Views:


Two models available, 30° to 350°C (662°F), & -12° to 140°C (284°F)

  • Wide temperature ranges: covers most critical test points
  • Fast: Nine minutes to reach temperature
  • Better accuracy: 0.4°C to 0.2°C, using comparison techniques
  • Light, compact, and portable
  • Interchangeable inserts: have one unit with multiple inserts to suite different sensor diameters
Base Price: $1,995.00


Product Code: QUICKCAL


Model Selection*:
  

   Product Description
 
  • OVERVIEW
  • SPECIFICATIONS
  • DOWNLOADS

Handheld Temperature Calibrator "Quick Cal" for Quick Calibration

Isotech’s Quick Cal range of testers are ideally suited to the less exacting applications where our larger, better specified blocks are unnecessary.

The Low Temperature Quick Cal covers the temperature range of -12°C to +140°C, making it ideal for the medical, sterilization and food industries. The High Temperature Quick Cal covers the temperature range from 30°C to 350°C.

Both have interchangeable inserts, so you don’t have to buy another calibrator each time you find a new diameter sensor to calibrate. The Low Temperature Quick Cal is particularly exciting since it can work from a small 12 volt battery, making it truly portable, alternatively a mains supply is available with 100V - 240V, 50/60Hz input. The Low Temperature Quick Cal comes complete with two inserts, one blank and one drilled 6.8mm diameter 120mm deep. Alternatively we can provide pre-drilled inserts to suit your needs. Each calibration well is 13mm diameter and 120mm deep.

The High Temperature Quick Cal has a single larger well and interchangeable inserts, 25mm in diameter by 120mm deep. Blank and other special drillings are available to order. The High Temperature Quick Cal is supplied 100-130 or 200-250V AC, 50/60Hz.
QuickCal 550 QuickCal 560
Temperature Range 30°C to 350°C (662°F) -12°C to 140°C (284°F)
Set Point Resolution 0.1°C over range 0.1°C over range
Accuracy T < 100°C = 0.4°C against display
T > 100°C = 1.0°C against display
±0.4°C against display with single probe in block
Stability ±0.05°C ±0.05°C
Heating Time From ambient to 350°C in 9 min From 0°C to 100°C in 9 min
Additional Time
For Best Stability
Typically in 5 minutes Typically in 5 minutes
Calibration Volume 25mm diameter by 120mm deep (2) 13mm diameter sleeves by 120mm deep
Standard Insert 8mm, (2) 6.5mm, (2) 4.5mm diameter by 115mm deep (1) 6.8mm diameter and (1) Blank sleeve 120mm deep
Power 120 VAC ±10% (50 / 60 Hz) 15 VDC or 120 VAC ±10% (50 / 60 Hz)
Overall Dimensions 65mm x 152mm x 175mm (HxWxD) 65mm x 152mm x 175mm (HxWxD)
Weight 3.3 lbs (1.5kg) 3.3 lbs (1.5kg)
Literature

Quick Cal Datasheet Quick Cal Manual Quick Cal Datasheet
Quick Cal Datasheet Quick Cal Manual Español